System to measure the angle position of a large rotating ring structure

Request Number N271088
Author Siemens
Need details
Description

Siemens Healthineers is looking for a cost optimized solution to measure the dynamic angle position of a CT gantry. The position information is used for image reconstruction and motion control.

 

Background

State-of-the-art CT gantrieshave a rotating ring that moves the measurement system around the patient. It is a continuous movement with rotation speeds up to 320 U/min during which an x-ray based scan of the patient is performed.

 

The current state-of-the-art is having markers on the Gantry, either directly within the Gantry structure or on a carrier structure that is sampled by optical or inductive sensors.

 

Key Success Criteria

We are seeking partners that can provide a solution meeting the following criteria
·      Fit to CT Gantry geometry with the inner (open) diameter of the ring is around 900mm; the outer diameter is around 1800mm

·      Measure angle with a maximum absolute Deviation ±0.01 degree
·      Solution can scale to deliver orders of 1000 p.a.

 

Possible Approaches

Approaches that would drastically simplify the sensor are of special interest. Approaches that can be used to control a direct drive of the ring would be of additional interest.

 

Approaches not of interest

 
Preferred Collaboration types
  • Technology Licensing
  • Supply Agreement
  • To Be Negotiated
Items to be submitted

We are looking for concise, non-confidential proposals. The proposal should:

  • Describe the technical approach
  • Include a description of the responding organisation or team and their experience.

To Respond, please use the Respond Now button located at the top of this page.

 

Please note that only non-confidential information can be accepted. By submitting a response you represent that the response does not contain and will not be deemed to contain any confidential information of any kind whatsoever.  By submitting a response, you acknowledge that Siemens, in its sole discretion, may select or reject a response or any portion thereof.

 

Your NineSights community profile should contain all required information in order to provide us with appropriate contact information for your responses. For questions about NineSights privacy and security, please feel free to post in our Community Help forums.

Point of Contact
Stephen Clulow
Area of Interest
  • Electronics & Electronic Components
  • Medical Products & Services
Organization
Siemens
Due Date
December 9, 2016