Patient weight measurement system to be integrated into medical equipment / patient tables

Request Number

NG_2017_0013_N_012
Point of Contact
Need was closed on
Aug 8
Description

Siemens Healthcare is seeking a subsystem solution to be integrated into medical equipment (for example patient tables) which can measure or deduce the patient weight by either (1) complying with norm MID (Measurement Instruments Directive 2014/32/EU, “scale for medical purpose”)  or (2) not using gravitational effects.

 

Background

Integrated measuring of the patient weight can support the medical staff and improve the workflow in the hospital.

 

Key success Criteria
  • Measuring patient weight from 50kg to 150kg
  • Complying with regulatory requirements for medical scales (Measurement Instruments Directive 2014/32/EU, “scale for medical purpose”)
  • Tolerance: +/- 2 kg on a weighted weight of 50 to 150 kg

    OR

 

  • Methods NOT based on gravimetric measurements (preferred)
  • Communication interface using standard protocol (preferable CANopen)
  • No display required

 

Possible Approaches
  • Measuring device could be mounted under/on/on top of or integrated into the medical devices.
  • Weight can be deducted by a system camera.
  • Weight information can be deducted by combining several sources of patient information.

 

Approaches not of Interest

Solutions not supporting any standard communication interface (information output via display only is not sufficient)

 

Due Date
Jun 29
Items to be submitted

We are looking for concise, non-confidential proposals. The proposal should:

  • Describe the technical approach
  • Include a description of the responding organization or team and their experience.

To respond, please use the Respond button located to the right of this page.

 

Please note that only non-confidential information can be accepted. By submitting a response you represent that the response does not contain and will not be deemed to contain any confidential information of any kind whatsoever.  By submitting a response, you acknowledge that Siemens, in its sole discretion, may select or reject a response or any portion thereof.

 